@article{oai:ous.repo.nii.ac.jp:00001102, author = {猶原, 順 and Naohara, Jun and 山下, 栄次 and Yamashita, Eiji}, journal = {岡山理科大学紀要. A, 自然科学, Bulletin of Okayama University of Science. A, Natural Sciences}, month = {Mar}, note = {P(論文), The trace elements in water reference materials (NIST 1643c and NRC SLRS-2) were determined by inductively coupled plasma mass spactrometry (ICP-MS). Calibration was accomplished by using blank and standard solution, and finally the standard solution was run again to allow a correction for intensity drift. The difference between the concentration of trace elements in water reference materials and the concentration of trace elements in diluted water reference materials with MILLI-Q water was compared. In NIST 1643c reference materials, the concentrations of Be, B, Na, Mg, Al, V, Cr, Mn, Fe, Co, Ni, As, Rb, Sr, Mo, Ba and Pb in diluted water reference material were close to the certified concentration. In NRC SLRS-2 reference material, the concentration of Na, Mg, Ca, Cr, Mn, Co, Zn, Mo and Sb in diluted water reference material were close to the cortified concentration.}, pages = {131--140}, title = {ICP-MSによる環境試料中の痕跡元素の測定 : (3)ICP-MSによる水標準試料中の痕跡元素の測定}, volume = {30}, year = {1995}, yomi = {ナオハラ, ジュン and ヤマシタ, エイジ} }