@article{oai:ous.repo.nii.ac.jp:00001060, author = {猶原, 順 and Naohara, Jun and 佐藤, 昭彦 and Satou, Akihiko and 川崎, 尚芳[他] and Kawasaki, Naoyoshi}, journal = {岡山理科大学紀要. A, 自然科学, Bulletin of Okayama University of Science. A, Natural Sciences}, month = {Mar}, note = {P(論文), Fast and efficient sample digestion is achieved by a microwave-nitric acid procedure. The ICP-MS analysis is performed with single standard calibration. The optimum condition of microwave digestion is determined. Dried samples of 0.1g were digested into the Teflon cup of high-pressure microwave acid-digestion bomb for 20-30 seconds. For all elements except Ca and Cu, good agreement exists between ICP -MS and certified values.}, pages = {139--148}, title = {ICP-MSによる環境試料中の痕跡元素の測定 (2)ICP-MSによる環境標準試料(NIES茶葉)中の痕跡元素の測定}, volume = {29}, year = {1994}, yomi = {ナオハラ, ジュン and サトウ, アキヒコ and カワサキ, ナオヨシ} }